Lab Research Products (LRP) - LRP-HC-FFP-14 - 14 Cu. Ft. Flammable Storage Freezer Basic Series with Natural Refrigerant

Features
Freezer models feature a microprocessor temperature controller and display, allow for precise temperature control, verification and recovery versus mechanical style thermostats.
Environmentally friendly, natural hydrocarbon (HC) refrigerants (not containing hydrofluorocarbons) which vastly reduce global-warming potential (GWP) while meeting new EPA/SNAP mandates and UL or ETL, ASHRAE compliance guidelines.
Please Note: For manual defrost freezers, periodic manual defrosting is required. Freestanding models require 4″ of clearance around back and sides of unit for proper ventilation.
Specifications
Overview
-
Nominal Height:73" (185.42 cm)
-
Nominal Width:23.63" (60.02 cm)
-
Nominal Length:24.38" (61.93 cm)
-
Internal Width:19" (48.26 cm)
-
Internal Length:17.5" (44.45 cm)
-
Capacity:14
General Description and Application
-
Door:Right Hinged
-
Shelves:6 Fixed
-
Freezer Compartments:Non-applicable
-
Baskets:Non-applicable
-
Mounting:Leveling Legs
-
Interior Lighting:Non-applicable
-
Airflow Management:Non-applicable
-
External Probe Access:Rear wall port (3/4" dia.)
-
Insulation:Cabinet is foamed-in-place with EPA compliant high density urethane foam
-
Exterior Materials:White powder coated steel
-
Access Control:Keyed Door Lock
-
Shipping Weight (lbs):190
-
Rated Amperage:3 Amps
-
Power Plug/Power Cord:NEMA 5-15 Plug, 8-10 ft. typical, conforms to UL471 requirements
-
Facility Electrical Requirement:115V, 60 Hz
-
Agency Listing and Certification:ETL, C-ETL listed, Energy Star Certified
Refrigeration System
-
Refrigerant:R290
-
Defrost:Manual
Controller, Configuration, Alarms and Monitoring
-
Controller Technology:Microprocessor temperature controller with digital temperature display, °C/°F switchable
-
Display Technology:Standard Digital Temperature Display
-
Adjustable Temperature Range:-15°C to -25°C
-
External Alarm Connection:State switching remote alarm contacts
-
Alarms:Audible and visual high and low temperature alarms
-
Controller Probe:2 Probes; 1 in air and 1 in sample bottle
-
Disclaimers:These units are NOT designed for use in environments where volatile/explosive conditions could potentially exist; Defrosting of the unit will need to be performed periodically to prevent severe ice build-up; Indoor use only, +18°C to +26°C (+65°F to +78°F), <70% RH
Downloads
-
LRP-HC-FFP-14 Data SheetDownload