FARRAR - 4002-LC - 4002-LC - Controlled Rate Freezer, Air-Cooled

Features
Controlled Freeze/Thaw
The 4000 Series Controlled Rate Chamber offers uniformity and repeatability in rapid, controlled freezing and thawing applications. Our technology uses forced air circulation to provide efficient pulldown from ambient to the desired temperature. Forced air circulation allows the rapid pulldown process to extend to the fluid core, avoiding false freezing points and optimizing the freezing process (1,000 CFM air circulation).
This unique, convection-based, controlled freezing and thawing process eliminates uncertainty and allows temperature-specific conditioning to satisfy processing and shipping protocols for various pharmaceuticals. Our 4000 Series Controlled Rate Chamber units can meet your freeze/thaw application needs from +40°C to -80°C.
Specifications
Overview
-
Nominal Height:80" (203.2 cm)
-
Nominal Width:75" (190.5 cm)
-
Nominal Length:38" (96.52 cm)
-
Internal Height:43" (109.22 cm)
-
Internal Width:34" (86.36 cm)
-
Internal Length:27.5" (69.85 cm)
Performance Data
-
Pull Down Empty Chamber:< 38 minutes from +25°C to -80°C
-
100L Load +25°C to -80°C:< 12 Hours
-
Uniformity Air Temperature:± 2.0°C
-
Uniformity Product Temperature:± 1.0°C
-
Net Weight:1,010 lbs. (459 kg)
-
Shipping Weight:1,597 lbs. (726 kg)
Refrigeration System
-
Heat Transfer:Convection, Air Flow Evaporator
-
High Stage:R-449A
-
Low Stage:R-508B
-
Defrost:Defrost-Manual Start/Automatic Complete
-
Heat of Rejection/Air-Cooled:38,000 BTU/hr (11 Kw/hr)
Controller
-
Controller:IDEC PLC with 12" Display and 4 outputs (General Alarm, Profile Running, Profile Complete, Door Open), 0-10VDC outputs for chamber probe and optional second probe
-
Sensor:RTD PT100 Din A
-
Dry Contacts:Standard (C, NO, NC)
Useful Links
-
FARRAR LC Controlled Rate Chamber Tech Data SheetVisit Link