American Biotech Supply - ABT-VS-36S - 36 Cu. Ft. Premier Solid Door Laboratory Refrigerator with Variable Speed Compressor

Features
They feature precise microprocessor temperature control and variable speed compressor technology. Enjoy low energy consumption and natural hydrocarbon refrigerants for cost-savings and energy efficiency. Security and monitoring features include a keyed door lock and comprehensive alarms with remote alarm contacts. The cabinet is built with durable powder-coated steel, a self-closing solid door, safety-shielded LED lighting, and adjustable interior shelving to enhance your workflow. This model delivers optimal temperature stability and includes a 2-year parts and labor warranty with a generous 7-year warranty on the compressor. These premier refrigerators are designed to increase performance and efficiency in your lab.
Specifications
Overview
-
Nominal Height:81.75" (207.65 cm)
-
Nominal Width:39.625" (100.65 cm)
-
Nominal Length:34.75" (88.27 cm)
-
Internal Height:48.75" (123.83 cm)
-
Internal Width:34.75" (88.27 cm)
-
Internal Length:26.5" (67.31 cm)
-
Capacity:36
General Description and Application
-
Door:Two (2) swing solid doors, self-closing, one right hinged, one left hinged, magnetic door gasket
-
Shelves:Eight (8) adjustable shelves (adjustable in 1/2" increments)
-
Mounting and Installation:Swivel casters - locking front casters
-
Interior Lighting:LED interior lights are safety shielded and switch controlled
-
Airflow Management:Forced draft air circulation, patented air flow technology
-
External Probe Access:3/4" probe access port on rear wall
-
Insulation:Cabinet is foamed-in-place with EPA compliant high-density urethane foam
-
Exterior Materials:White powder coated steel
-
Access Control:Keyed door lock
-
General Warranty:Two (2) year parts and labor warranty
-
Compressor Warranty:Seven (7) year compressor warranty
-
Product Weight (lbs):345
-
Shipping Weight (lbs):395
-
Rated Amperage:3 Amps
-
Power Plug/Power Cord:NEMA 5-15P plug
-
Facility Electrical Requirement:110 - 120V AC, 15A breaker, NEMA 5-15 receptacle
-
Agency Listing and Certification:ETL, C-ETL listed certified to UL471 standard. Meets or exceeds Energy Star v2.0 for Laboratory Refrigerators
Performance
-
Maximum Temperature Variation (Cabinet Air):+/- 1.5°C
-
Stability (Simulator Ballast):+/- 0.3°C
-
Adjustable Temperature Range:2°C to 15°C
-
Energy Consumption (KWh/day):<2.2 KWh/day *meets ENERGY STAR® v2.0 for Laboratory Refrigerators
-
Average Heat Rejection (BTU/hr):<570 BTU/h
-
Noise Pressure Level (dBA):40 dBA or less installed
-
Pull Down Time to Nominal Operating Temp:< 40 min (empty)
-
Performance Note:Max temperature variation based on an empty chamber at steady state using 15 bare (unweighted) thermocouple probes. Product loading, door openings, and ambient conditions will affect performance. Stability of simulator ballast based on bottle probe provided with the unit located center of empty chamber at steady state, no door openings.
Refrigeration System
-
Compressor:Hermetic, variable speed
-
Refrigerant:EPA SNAP compliant R600a Isobutane (Hydrocarbon natural refrigerant)
-
Condenser:Tube and grid, high efficiency fan forced air
-
Evaporator:Fin and tube, high efficiency fan forced air
-
Defrost:Cycle Defrost
Downloads
-
ABT-VS-36S DataSheetDownload