Amerex - AT-HRH-110 - Front-Loading Autoclave with Fully Automatic Drying Cycle Using Heat and Vacuum

Features
- Front-loading design
- Fully automatic operation from start to finish with or without drying.
- One-touch door closing lever opens and closes the door easily, surely and safely.
- Pre-vacuum cycle - vacuum pump removes air from the chamber when sterilizing solids such as glassware and pipette tips, thereby eliminating cold spots, and ensuring faster and more efficient sterilization.
- Drying by heat and vacuum increases drying efficiency and reduces cycle times.
- Additional drying mode for complete drying without having to run the sterilization cycle again.
- Bacteria filter reduces re-contamination.
- Lid safety interlock prevents lid opening during a cycle, when the chamber is pressurized, or when the chamber temperature is above 97°C at which super-heated liquids might boil over when disturbed.
- Door-closure sensor prevents a cycle from starting unless the door is properly closed.
- HRH-110 size: 110 liters, 16.5" diameter, 31.3" depth.
- Options: 1) Load sensor, square type wire basket, wire basket with solid bottom, dressing case, 2) Data logger, and 3) Digital printer.
Sterilization and Operation Modes
1. For solids - pre-vacuum → sterilization → drying
2. For solids - pre-vacuum → sterilization → fast exhaust
3. For liquids - sterilization → slow exhaust
4. For drying - additional drying after mode 1 → end
Safety Devices
Pressure relief valve, circuit breaker with GFCI, vacuum pump breaker, low-water power cut-off, insufficient water in the exhaust bottle warming lamp, error messages for:
1) wire breakage in temperature sensor
2) over-heating
3) over-cooling
4) over-pressure
5) door is not properly locked.
Specifications
Overview
-
Nominal Height:54.5" (138.43 cm)
-
Nominal Width:27.2" (69.09 cm)
-
Nominal Length:45.3" (115.06 cm)
Technical Data
-
Self-Contained Design:Does not require in-house steam, plumbing or a floor drain hole for installation
-
Sterilization Temperature:105°C - 135°C (programmable)
-
Temperature Display Range:5°C - 137°C
-
Sterilization Timer:1 ~ 60 min (remaining time displayed)
-
Drying Timer:1 ~ 120 min (remaining time displayed)
-
Drying Method:Vacuum pump and heating
-
Chamber Water Source:Auto water filling from an internal 10L water tank
-
Maximum Allowable Pressure (Gauge Pressure):0.206 MPa (2.06 kg/cm2G)
-
Chamber Size (Diameter x Depth):110 liters, 16.5" x 31.3" (420 x 795 mm)
-
Chamber Material:Stainless steel SUS304
-
Pressure Gauge:Analog display
-
Standard Accessories:Drain hose, water supply hose, caster stopper
-
Power Source/Consumption:AC220V, 50/60Hz, 1 phase, 3.2kW
-
Net Weight (Approx.):462 lb, (210 kg)
-
Warrenty:1 year