American Biotech Supply - ABT-HC-EFP-14 - 14 Cu. Ft. Capacity Standard Hazardous Location Freezer

Features
Our Hazardous Location Freezers meet NFPA and OSHA guidelines for Hazardous Location, Class 1, Division II, Groups A,B,C & D guidelines. These units are engineered for use in environments where volatile / explosive conditions could potentially exist. For storage of flammable or volatile materials within typical laboratory or non-volatile environments, see Flammable Material Storage units.
Utilizes ATEX and non-sparking components for safety. Models feature a microprocessor temperature controller, allowing for precise temperature control, recovery and maintenance versus mechanical style thermostats.
Models utilize environmentally friendly, natural hydrocarbon (HC) refrigerants (not containing hydrofluorocarbons) which vastly reduce global-warming potential (GWP) while meeting new EPA/SNAP mandates and UL, ASHRAE compliance guidelines.
Please Note: Not equipped with plug; must be hardwired in conduit. Periodic Manual Defrosting is required.
Specifications
Overview
-
Nominal Height:73" (185.42 cm)
-
Nominal Width:23.63" (60.02 cm)
-
Nominal Length:24.38" (61.93 cm)
-
Internal Width:19" (48.26 cm)
-
Internal Length:17.5" (44.45 cm)
-
Capacity:14
General Description and Application
-
Door:Right Hinged
-
Shelves:6 Fixed
-
Freezer Compartments:Non-applicable
-
Baskets:Non-applicable
-
Mounting:Leveling Legs
-
Interior Lighting:Non-applicable
-
Airflow Management:Non-applicable
-
External Probe Access:Non-applicable
-
Insulation:Cabinet is foamed-in-place with EPA compliant high density urethane foam
-
Exterior Materials:White powder coated steel
-
Access Control:Keyed door lock
-
General Warranty:Two (2) year parts and labor warranty
-
Warranty Disclaimer:Products require 4" of clearance around back and sides of the unit for proper ventilation - improper installation will void the warranty and will lead to temperature maintenance issues, overheating and possible failure
-
Product Weight (lbs):170
-
Shipping Weight (lbs):190
-
Rated Amperage:3 Amps
-
Power Plug/Power Cord:NEMA 5-15 Plug, conforms to UL471 requirements
-
Facility Electrical Requirement:115V, 60 Hz
-
Agency Listing And Certification:ETL, C-ETL listed and certified to UL471 standard, hydrocarbon refrigerant safety, Energy Star Certified. Units meet NFPA and OSHA guidelines for Hazardous Location, Class 1, Division II, Groups A,B,C & D guidelines.
Refrigeration System
-
Refrigerant:R290
-
Defrost:Manual
Controller, Configuration, Alarms and Monitoring
-
Controller Technology:Microprocessor temperature controller with digital temperature display, °C/°F switchable
-
Display Technology :Standard Digital Temperature Display
-
Adjustable Temperature Range:-15°C to -25°C
-
External Alarm Connection:State switching remote alarm contacts
-
Alarms:Audible and visual high and low temperature alarms
-
Controller Probe:Non-applicable
-
Digital Communication:Non-applicable
-
Data Transfer:Non-applicable
Downloads
-
ABT-HC-EFP-14 Data SheetDownload