ESCO - STL/SC-M3.5 - ISO Class 5 Soft Wall Cleanroom

Product Range
SKU | Name |
---|---|
STL/SC-887-M3.5 | ESCO - STL/SC-887-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-10127-M3.5 | ESCO - STL/SC-10127-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-8207-M3.5 | ESCO - STL/SC-8207-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-6167-M3.5 | ESCO - STL/SC-6167-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-8167-M3.5 | ESCO - STL/SC-8167-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-8107-M3.5 | ESCO - STL/SC-8107-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-8127-M3.5 | ESCO - STL/SC-8127-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-6207-M3.5 | ESCO - STL/SC-6207-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-12127-M3.5 | ESCO - STL/SC-12127-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-6127-M3.5 | ESCO - STL/SC-6127-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-687-M3.5 | ESCO - STL/SC-687-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-487-M3.5 | ESCO - STL/SC-487-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
STL/SC-447-M3.5 | ESCO - STL/SC-447-M3.5 - Soft Capsule® Soft Wall Cleanroom ISO Class 5* (US Federal Standard Class 100) |
Features
- Cleanroom ceiling grid enables flexible placement of fan filter units. Non-particulate shedding powder coated steel panels are provided to cover empty bays.
- Provides ISO Class 5, 6, or 7 environment (depends on client specifications) via high quality HEPA filters utilizing an improved mini-pleated separation technique to maximize surface area, improve efficiency, and extend the filter life.
- Centrally mounted control panel enables rapid start-up of fan filter units and lighting.
- The energy efficient external rotor motor design reduces operating costs and has extremely low noise and vibration levels, improving the working conditions in the cleanroom.
- Isostat™ vinyl curtains isolate the cleanroom from the ambient environment, while allowing easy passage of materials and personnel in and out of the cleanroom area.
- Cleanroom grade construction utilizing fully-welded, reinforced, steel tubular sections. Structure is finished with an abrasion-resistant oven-baked powder coat.
- Entire structure is free-standing and does not require any suspending ceiling supports.
- Built-in warm white, 5000k lighting provides excellent illumination of the cleanroom, and reduces operator fatigue.
Applications
- Pharmaceuticals, Grade A filling suites
- Medical devices, plastic injection moulding
- Electronics assembly
- Contact lens packaging
- Hospital pharmacy (USP 797)
- Biotechnology
- Nanotechnology
- Aerospace
- Quality control
- Food industries
Principle
- Room air is taken in from the top of the cleanroom through a disposable pre-filter with 85% arrestance; this serves to trap larger particles and increase the life of the main filter.
- Air is forced evenly across the HEPA filter(s); the result is a stream of clean laminar air within the work zone; this dilutes and flushes all airborne contaminants from the interior.
- A nominal filter face velocity of 0.45 m/s or 90 fpm ensures that there is a sufficient number of air changes within the cleanroom in order to maintain cleanliness.
- The purified air travels downward within the interior in a vertical, unidirectional stream, and leaves the cleanroom close to floor level at the perimeter.
- The cleanroom is positively pressurized to prevent ingress of airborne contaminants from the external environment.
Options
- All stainless steel construction
- Leveling feet
- Complete factory test and certification, including airflow, filter integrity, particle counting, light and noise tests
- Extensions for existing Soft Capsule cleanrooms