Baker - NCB605 - 6' NCB™ e3 Class II Type B1 Biological Safety Cabinet

Features
Developed to exceed the National Cancer Institute's Expectations and Yours
The Baker NCB™ e3 biological safety cabinet exceeds specifications for the NCI-1 cabinet developed by the National Cancer Institute, with unique design features proven to enhance containment while inhibiting cross contamination and exposure to chemicals, vapors and gases. Patented under the work surface, HEPA filters and the direct exhaust airflow system combined, increases protection from particulates, preventing the buildup of chemical vapors and gases. All exhaust air is removed directly from the work area (direct exhaust) into a building's separate exhaust system. Vapors and gases emitted from vessels or work behind the air split (approximately halfway back from the cabinet front) are removed and not recirculated.
HEPA filters located below the work surface assure that all positive pressure areas are free of particulate contamination. Additionally, all positive pressure plenums are completely surrounded by negative pressure. Recirculated air is HEPA-filtered immediately below the work surface before it is passed through a supply downflow filter above the work area.
Energy Efficient Engineered.
The NCB™ e3 utilizes StediFLOW™, a self-adjusting motor technology that requires less energy, extends filter life, and reduces heat output to the lab without sacrificing performance.
Specifications
Overview
-
Nominal Height:99.75" (253.37 cm)
-
Nominal Width:77.63" (197.18 cm)
-
Nominal Length:34" (86.36 cm)
-
Internal Height:26" (66.04 cm)
-
Internal Width:70" (177.8 cm)
-
Internal Length:18.88" (47.96 cm)
Technical Specifications
-
Size:6'
-
Usable Workspace (w x d):70.0" x 17.06"
-
Exterior Dimensions (w x d x h [Min to Max]):77.63" x 34.0" x [91.25" to 99.75"]
-
Cabinet Weight:1,200 lbs
-
Shipping Weight:1,350 lbs
-
Working Access Opening Height (8" Standard):8" / 12"
-
Opening Max:17.5"
-
Operating Amperage:5.5 Amps
-
Power Consumption:633 Watts
-
Heat Generation:2,159 BTU/hr
-
Electrical Service Requirement:115V AC, 20 A, 60 Hz, 1Ø, 16 A max usable current
-
Noise, Exhaust & Cabinet:63 dBA
Drawings
-
NCB605 - Standard Detail DrawingVisit Link
Useful Links
-
NCBe3 Operators ManualVisit Link
-
FAQ Regarding UV Lights and ViewscreenVisit Link
-
Lab Positive/Negative Balancing ConditionsVisit Link
-
New Stedivolt Installation InstructionsVisit Link
-
Use of Flames in Biological Safety CabinetsVisit Link