Baker - BCG601-115V - 6' BioChemGARD® e3 NSF Class II Type B2 Biosafety Cabinet, 115V

Features
The most energy efficient, comfortable and safe B2 cabinet in the industry
The BioChemGARD® e3 is a breakthrough in total-exhaust biological safety cabinets. Engineered specifically for laboratories that need containment and removal of vapors, mists and particulates, the BioChemGARD® e3 has a revolutionary airflow system that increases users' comfort and productivity while decreasing the overall operating costs.
Unique Airflow Design
Unique airflow design operates with a reduced exhaust airflow and resistance, which minimizes noise and reduces the volume of conditioned air that is exhausted from the laboratory. To find out more about energy efficiency, read our white paper.
ExchangeSAFE™ System
The motor/blower design provides up to a 420% reserve for a very long filter life, resulting in less maintenance. With the exclusive ExchangeSAFE™ (sealed access filter exchange system) located directly under the work surface, improves access making the process quiet with reduced exposure to containments.
Self-Adjusting Airflow
The StediFLOW™ VFD motor controller maintains precise airflow by self-adjusting for filter loading. Also controls the ReadySAFE™ low-flow operating mode that preserves containment conditions when the BSC is not in use and substantially reduces energy consumption.
Optional Fume Hood Package
Meets the UL requirements for Electrical/Mechanical Safety (UL 61010-1) and the UL requirements for Material Flammability and Effectiveness of Airflow Characteristics (UL 1805). The cabinet design has been tested in accordance with ANSI/ASHRAE 110-1995. Usage should be determined by a professional industrial hygienist.
A Fully Integrated Pharmacy Compounding Validation Software Package
To optimize the validation and documentation of sterile compounding procedures, Baker integrates with a variety of IV workflow management systems as well as Pharmacy Verification Software.
- Built-in Fully Integrated IV Prep Software keeps the compounding environment clean with no wires or devices within the hood. Compounding space is not altered.
- Uses a powerful 5-megapixel imaging system that closely focuses on the names, numbers, labels, color, and other details of drugs and syringes.
- Includes image recognition and barcode scanning features.
- Storage and retrieval image system allows documentation of the compounding process for quality purposes.
- Easily accessed from desktops, laptops, tablets, and mobile devices.
- Compounding status dashboard is automatically updated.
- Scalable and modular system is flexible and adaptable to the customer environment.
Specifications
Overview
-
Nominal Height:94.5" (240.03 cm)
-
Nominal Width:77.88" (197.82 cm)
-
Nominal Length:33.6" (85.34 cm)
-
Internal Height:27.63" (70.18 cm)
-
Internal Width:70" (177.8 cm)
-
Internal Length:22.88" (58.12 cm)
Technical Specifications
-
Nominal Size:6'
-
Usable Workspace (w x d x h):68.0'' x 17.63'' x 27.63''
-
Exterior Dimensions (w x d x h [Min to Max]):77.88'' x 33.63'' x [86.0'' to 94.5'']
-
Cabinet Weight:850 lbs
-
Shipping Weight:1020 lbs
-
Sash Height:8"
-
Opening Max:20"
-
Operating Amperage - Normal Operating Mode:1.4 Amps
-
Operating Amperage - ReadySAFE Mode:0.3 Amps
-
Power Consumption - Normal Operating Mode:161 Watts
-
Power Consumption - ReadySAFE Mode:35 Watts
-
Heat Generation (100% Exhausted) - Normal Operating Mode:549 BTU/hr
-
Heat Generation (100% Exhausted) - ReadySAFE Mode:118 BTU/hr
-
Electrical Service Requirements:115V AC, 20 A, 60 Hz
-
Noise, Exhaust & Cabinet (NSF International):60 dBA
-
Noise, Cabinet:<50 dBA
Downloads
Drawings
-
BioChemGARD e3 BCG 601 with Air Tight Damper DrawingVisit Link
Useful Links
-
BioChemGARD® e3 Operator's ManualVisit Link
-
BioChemGARD Accessories & UpgradesVisit Link
-
BioChemGARD e3 Purchasing Specs (US115V)Visit Link
-
AIA Master SpecificationVisit Link