American Biotech Supply - ABT-ERCS-30 - 30 Cu. Ft. General Purpose Hazardous Location Refrigerator/Freezer Combination

Features
These units are engineered for use in environments where volatile / explosive conditions could potentially exist. They are not intended for common lab / room environments that are not designed to be explosion proof (See Flammable Material units).
Units are designed with no internal electrical components and all external components are sealed per NFPA and OSHA guidelines.
Units must be hard wired in conduit to 120VAC power source. Power cords with plugs are NOT ACCEPTABLE.
Details
- Three adjustable refrigerator shelves
- Keyed door lock
- Magnetic door gasket for positive seal
- Stainless steel door handle
- Leveling legs
- CFC-free refrigerant (R404a), foam insulation and packaging
- Not equipped with line cord and plug; must be hardwired in conduit to 120 VAC power source
- One year parts and labor warranty
- UL/C-UL listed
- Meet NFPA and OSHA guidelines for hazardous locations, Class 1, Division II, Groups C & D.
- These units are not intended for use in general clinical or research laboratories.
Temp Controller
- Mechanical thermostat sealed in UL listed explosion proof box
- Adjustable operating temperature range: Refrigerator: 1°C to 10°C, Freezer: -15°C to -25°C
Specifications
Overview
-
Nominal Height:77.5" (196.85 cm)
-
Nominal Width:35.75" (90.81 cm)
-
Nominal Length:36.25" (92.08 cm)
-
Capacity:30
Technical Data - Freezer
-
Defrost:Manual
-
Door:1 Solid
-
Int Door:1
-
Shelves:0
-
Interior Dim. W":29
-
Interior Dim. D":22.25
-
Interior Dim. H":12
-
Refrigerant:R404a
-
H.P.:1/3
-
Amps:6
-
Weight:348 lbs.
Technical Data - Refrigerator
-
Defrost:Manual
-
Door:1 Solid
-
Int Door:0
-
Shelves:3 Adj.
-
Interior Dim. W":32
-
Interior Dim. D":22.5
-
Interior Dim. H":50.25
-
Refrigerant:R404a
-
H.P.:1/3
-
Amps:6
-
Weight:348 lbs.