Memmert - HPP 110eco 100-240Volt - Constant Climate Chamber

Memmert - HPP 110eco 100-240Volt - Constant Climate Chamber

Memmert - HPP 110eco 100-240Volt - Constant Climate Chamber

Contact Brent for Assistance

Need help right away? Give our customer service team a call for immediate attention

(510) 909-5846
Brent Kolhede

Ask About This Item

Add the product to a quote to receive more information.

Ask an Expert

Memmert Climate Chamber HPP 110eco, 100-240volt Temp Range 0°C-70°C Humidity Range 10-90%RH D8-40mm silicone entry port with stopper rear wall.

Features

Maximum energy efficiency in continuous operation - tailor-made for stability studies according to ICH guidelines, stability tests for cosmetics and foods as well as environmental testing and material testing. HPP110 Memmert constant climate chambers guarantee the perfect atmosphere for climate and temperature tests. Our HPP110eco constant climate chamber is quiet, space-saving, environmentally friendly, and extremely economical.

 

  • Stainless Steel: Scratch resistant, hygienic and durable.
  • 3 Year Warranty: A limited 3 year warranty is included with the option to purchase an extra year.
  • AtmoCONTROL: Create and monitor programs.
  • TwinDISPLAY: Twice the features of SingleDISPLAY. Monitor programs easily via Ethernet or USB.

Specifications

Overview

  • Nominal Height:
    34" (86.36 cm)
  • Nominal Width:
    29.3" (74.42 cm)
  • Nominal Length:
    21.8" (55.37 cm)
  • Internal Height:
    18.9" (48.01 cm)
  • Internal Width:
    22" (55.88 cm)
  • Internal Length:
    15.7" (39.88 cm)

Temperature - Humidity

  • Setting Temperature Range without Light, with Humidity:
    +5 to +70°C
  • Setting Temperature Range with Light, with Humidity:
    +5 to +70°C
  • Setting Temperature Range with Light, without Humidity:
    0 to +70°C
  • Setting Temperature Range without Light, without Humidity:
    0 to +70°C
  • Working Temperature Range with Light, without or with Humidity:
    15 to +40°C
  • Working Temperature Range without Light, with Humidity:
    +5 (at least 20 below ambient temperature) to +70°C
  • Working Temperature Range without Light, without Humidity:
    0 (at least 20 below ambient temperature) to +70°C
  • Setting Accuracy Temperature:
    0.1°C
  • Temperature Sensor:
    2 Pt100 sensors DIN Class A in 4-wire-circuit
  • Setting Range Humidity:
    10 - 90 % rh
  • Humidity:
    Humidity supply with distilled water from external tank by self-priming pump
  • Setting Accuracy Humidity:
    0.5%
  • Humidification:
    By hot steam generator
  • Dehumidification:
    By cold trap using the Peltier technology

Interior - Exterior - Control Technology

  • Stainless Steel Interior - Volume:
    108 L
  • Stainless Steel Interior - Max. Number of Internals:
    5
  • Stainless Steel Interior - Max. Loading per Internal:
    20 kg
  • Stainless Steel Interior - Max. Loading of Chamber:
    150 kg
  • Textured Stainless Steel Casing - Housing:
    Rear zinc-plated steel
  • Control Technology - ControlCOCKPIT:
    TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays
  • Control Technology - Language Setting:
    German, English, Spanish, French, Polish, Czech, Hungarian, Chinese
  • Control Technology - Adjustable Parameters:
    Temperature (Celsius or Fahrenheit), programme time, time zones, summertime/wintertime
  • Control Technology - Timer:
    Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
  • Control Technology - Function SetpointWAIT:
    The process time does not start until the set temperature is reached
  • Control Technology - Calibration:
    Three freely selectable temperature values

Safety - Standard Equipment - Electrical Data

  • Safety - Autodiagnostic System:
    Integral fault diagnostics for temperature and humidity control
  • Safety - Alarm:
    Visual and acoustic
  • Standard Equipment - Standard Works Calibration Certificate:
    +25°C / 60 % rh, +40°C / 75 % rh
  • Standard Equipment - Door:
    Fully insulated stainless steel doors with 2-point locking (compression door lock)
  • Standard Equipment - Internals:
    2 stainless steel grid(s), electropolished
  • Standard Equipment - Standard Accessories:
    Water tank including connection hose
  • Electrical Data - Electrical Load :
    230 V (±10%), 50/60 Hz, Approx. 420 W
  • Electrical Data - Electrical Load:
    115 V (±10%), 50/60 Hz, Approx. 420 W

Downloads

Useful Links

  • Data Sheet - Constant Climate Chamber HPP110eco
    Visit Link