FARRAR - Model 4000 Series Controlled Rate Chamber

Features
Our Model 4000 series rate chamber, purpose-built for bioprocessing applications, greatly reduces freeze/thaw conditioning times, measured in hours, instead of days or weeks.
This precision controlled rapid freeze/thaw chamber helps protect product quality by enabling fast, uniform bulk freezing and thawing of protein or products in a variety of containers of various sizes, (bags, bottles, etc.) including single use system (SUS), polycarbonate vessels or polypropylene vessels.
Our high-performing Controlled Rate Chamber model 4002/05, (air-cooled) or 4102/05, (water-cooled) conducts rapid freeze/thaw for +40°C to -80°C temperature needs. FARRAR's refrigeration experts help customers plan their process and provide performance data to ensure requirements are met.
- Wide-range temperature capability for nearly any application (+40°C to -80°C)
- Air or water-cooled options available
- Uniform, repeatable results to help ensure quality control
- Cycle start/stop with +/-1°C product temperature tolerance
- Sizable and powerful cooling capacity achieves steady state in 38min (empty chamber) and is capable of freezing 100L to -80°C in 12h
Specifications
Overview
-
Nominal Height:80" (203.2 cm)
-
Nominal Width:75" (190.5 cm)
-
Nominal Length:38" (96.52 cm)
-
Internal Height:43" (109.22 cm)
-
Internal Width:34" (86.36 cm)
-
Internal Length:27.5" (69.85 cm)
-
Capacity:23.3
Dimensions
-
Net Weight:1010 lbs. (459 kg)
-
Shipping Weight:1597 lbs. (726 kg)
Performance Data
-
Pull Down Empty Chamber:< 38 minutes from +25°C to -80°C
-
100L Load +25°C to -80°C:< 12 Hours
-
Uniformity Air Temperature:± 2.0°C
-
Uniformity Product Temperature:± 1.0°C
-
Controller:Watlow PID/Standard (NO, NC, C) with 4 outputs (General Alarm, Profile Running, Profile Complete, Door Open) 0-10VDC outputs for chamber probe and optional second probe
-
Sensor:RTD PT100 Din A
-
Dry Contacts:Standard (C, NO, NC)
Refrigeration System
-
Heat Transfer:Convection, Air Flow Evaporator
-
High Stage:R-449A
-
Low Stage:R-508B
-
Defrost:Defrost- Manual Start/ Automatic Complete
-
Heat of Rejection/ Air-cooled:38000 BTU/hr (11 Kw/hr)
-
Heat of Rejection/ Water-cooled:1700 BTU/hr (0.5 Kw/hr)