Caron - OUTL326-1 - Three Interior Outlets NEMA 5-15 (115V 60Hz)
Features
An optional interior electrical outlet is available to supply power to small interior appliances such as shakers or stirrers. It is not intended to power high current draw devices. Each outlet will handle up to 1.0 Amp if 115V or 0.5A if 230V. Incubators with three outlets will hold 3.0 amps total if 115V or 1.5A total if 230V. All outlets are resettable GFI protected.
This incubator is equipped with an internal hydrocarbon (HC) sensor. The status of the HC sensor is indicated by the green/yellow/red indicator lights near the outlets.
HC Sensor Status Indicator
LED Color Status
- Green: Sensor working and safe HC level, Energized*
- Yellow: Sensor is initializing, Not energized
- Red: Risky HC level detected, Not energized
- US outlet 115V, 60Hz
*If outlet is not energized, either the GFCI or fuse may have tripped.
R290 Refrigerant Units
DANGER - Flammable Refrigerant Used. Risk of fire or explosion
- No equipment that uses an open flame should be placed inside the refrigerator.
- Do not use instrumentation or equipment that incorporates potential ignition sources, e.g. open contact switching, brushed DC and AC motors, etc.
Specifications