Amerex - HV-25 - Portable, Top-Loading Autoclaves Without Drying

Amerex - HV-25 - Portable, Top-Loading Autoclaves Without Drying

Amerex - HV-25 - Portable, Top-Loading Autoclaves Without Drying

Contact Brent for Assistance

Need help right away? Give our customer service team a call for immediate attention

(510) 909-5846
Brent Kolhede

Ask About This Item

Add the product to a quote to receive more information.

Ask an Expert
Portable top-loading autolaves with advanced safety and user-friendly features.

Features

The 25-liter HV-25 and the 50-liter HV-50 autoclaves are self-contained, portable floor models mounted on caster wheels. They occupy minimum floor space, operate at 120V, and can fit conveniently into many locations in the laboratory. Their many uses include the sterilization of agar, media, waste, and glassware. The HV-50 model can accommodate a 10-L fermenter, a 15-L carboy, a 19” x 24” biohazard bag, nine 1-L, or twenty-one 500-mL bottles, or one hundred eighty 20-mm test tubes.

 

  • Space-saving design with lid opening upward.
  • Programmable auto-start for initiating a cycle up to 1 week later.
  • Dual-sensing lid interlock prevents lid opening when chamber is pressurized or when temperature exceeds 97°C.
  • Automatic air purging by a microprocessor prevents over-pressurization during a cycle.
  • One-touch lever to easily open and close chamber lid.
  • Programmable rate of steam exhaust.
  • Door-closure sensor provides a prompt and the cycle will not start unless the door is closed properly.
  • In-process display shows status and progress of sterilization cycle.
  • Agar sterilization mode for sterilizing agar and then automatically maintaining it at liquid temperature.
  • Unique accessory: Floating sensor to monitor load temperature.
  • Sterilization timer is activated only when load temperature reaches set temperature.
  • Agar melting mode for rapidly melting agar and maintaining it at liquid temperature, or for a hot and faster start-up.
  • Memory backup system retains user-programmed parameters even in case of power failure.

 

Safety Devices


Dual-sensing safety lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, overcurrent detector and circuit breaker.

Specifications

Overview

  • Nominal Height:
    37.4" (95.0 cm)
  • Nominal Width:
    19" (48.26 cm)
  • Nominal Length:
    18" (45.72 cm)
  • Capacity:
    25

Technical Data

  • Operating Range Temperature:
    105ºC to 126ºC
  • Display Range Temperature:
    128ºC
  • Warming Range Temperature:
    45ºC to 60ºC
  • Agar Dissolution:
    Heating at 60ºC to 100ºC for up to
  • Gauge Pressure - Operating Range:
    3 to 22 psi gauge (0.02 to 148 MPa)
  • Gauge Pressure - Maximum Allowable:
    27 psi (0.186MPa)
  • Timer - Sterilization:
    1 to 250 minutes
  • Timer - Delayed Auto-start:
    Program autoclave to start 1 minute to 7 days later
  • Overall Dimensions (W x D x H):
    (19 x 18 x 37.4) in, (48 x 46 x 95) cm
  • Chamber Dimensions (Diam. x Ht):
    25 liters, 9.4" x 21.6" (24 cm x 55 cm)
  • Source Power:
    120V, 50/60 Hz (220V/240V available)
  • Consumption Power:
    1500 watts (1500VA)
  • Warranty:
    Autoclaves are protected by a 12 month limited warranty