Amerex - HV-25 - Portable, Top-Loading Autoclaves Without Drying

Features
The 25-liter HV-25 autoclaves are self-contained, portable floor models mounted on caster wheels. It occupies minimum floor space, operate at 120V, and can fit conveniently into many locations in the laboratory. Its many uses include the sterilization of agar, media, waste, and glassware.
Space-Saving Design
Lid opens upward.
Programmable Auto-Start
For initiating a cycle up to 1 week later.
Dual-Sensing Lid Interlock
Prevents lid opening when chamber is pressurized or when temperature exceeds 97ºC.
Automatic Air Purging
By a microprocessor prevents over-pressurization during a cycle.
One-Touch Lever
To easily open and close chamber lid.
Safety Devices
Dual-sensing safety lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, overcurrent detector and circuit breaker.
Specifications
Overview
-
Nominal Height:37.4" (95.0 cm)
-
Nominal Width:19" (48.26 cm)
-
Nominal Length:18" (45.72 cm)
-
Capacity:25
Technical Data
-
Operating Range Temperature:105ºC to 126ºC
-
Display Range Temperature:128ºC
-
Warming Range Temperature:45ºC to 60ºC
-
Agar Dissolution:Heating at 60ºC to 100ºC for up to
-
Gauge Pressure - Operating Range:3 to 22 psi gauge (0.02 to 148 MPa)
-
Gauge Pressure - Maximum Allowable:27 psi (0.186MPa)
-
Timer - Sterilization:1 to 250 minutes
-
Timer - Delayed Auto-start:Program autoclave to start 1 minute to 7 days later
-
Overall Dimensions (W x D x H):(19 x 18 x 37.4) in, (48 x 46 x 95) cm
-
Chamber Dimensions (Diam. x Ht):25 liters, 9.4" x 21.6" (24 cm x 55 cm)
-
Source Power:120V, 50/60 Hz (220V/240V available)
-
Consumption Power:1500 watts (1500VA)
-
Warranty:Autoclaves are protected by a 12 month limited warranty
Downloads
-
HV-Series 25-50 BrochureDownload